Send us Fan Mail The Inorganic Ventures team answers listener questions focused on pharmaceutical elemental impurity analysis by ICP-MS. They explain why an indium internal standard becomes unstable in high-pH matrices due to insoluble hydroxide formation and how EDTA complexation can stabilize indium when used for iodine analysis. They discuss mercury washout problems caused by volatility, adsorption to sample introduction surfaces, memory effects, and redox/speciation, and emphasize proper ...